摘要原子力显微镜(Atomic force microscope, AFM)在轻敲模式下工作时,随着探针针尖与样品距离的逐渐减小,空气压膜阻尼的作用随之增大。为研究压膜阻尼对原子力显微镜振动系统的影响,分别使用无针尖探针和微球针尖探针进行扫频实验,并基于振动理论将该过程简化,得到了两种不同的振动模型的系统刚度。在考虑压膜阻尼作用影响后,将微球针尖振动系统模型进一步简化为一维振子模型,并对压膜阻尼的影响进行讨论。实验表明空气压膜阻尼模型对于探针样品在微尺度下的作用过程是准确合理的。该结果对原子力显微镜轻敲模式研究具有重要意义。
Abstract:In atomic force microscope (AFM) tapping mode, as the distance between the tip and the sample decreases, the effect of squeeze film air damping increases gradually. In order to explore the influence of squeeze film damping on the vibration system of AFM, the tune experiments were carried out using the tipless probe and the microsphere tip probe, and the process was simplified based on the vibration theory, in which the system stiffness of two different vibration models was obtained. Considering the influence of the squeeze film damping, the sphere tip vibration system model was further simplified to a one-dimensional oscillator model, and the influence of the squeeze film damping was discussed. Experiments have shown that the squeeze film damping model is accurate and reasonable for the action of probes and samples at micro-scale. The result is of great significance to the AFM tapping mode study.
柳世华,魏征,孙岩,刘运鸿. 压膜阻尼对原子力显微镜振动特性的影响研究[J]. 振动与冲击, 2020, 39(11): 177-183.
LIU Shihua, WEI Zheng, SUN Yan, LIU Yunhong. Study on the influence of squeeze-film damping on the vibration characteristics of atomic force microscopy. JOURNAL OF VIBRATION AND SHOCK, 2020, 39(11): 177-183.
[1] Binnig G, Quate C F, Gerber C. Atomic force microscope.[M].Oxford University Press, 1993:1412-1414
[2] 魏征,孙岩,王再冉,等. 轻敲模式下原子力显微镜的能量耗散[J]. 力学学报, 2017, 49(6):1301-1311.
[3] 张文明,闫寒,彭志科,等. 微纳机械谐振器能量耗散机理研究进展[J]. 科学通报, 2017(19):2077-2093.
[4] Dürig U. Conservative and dissipative interactions in dynamic force microscopy[J]. Surface & Interface Analysis, 2015, 27(5‐6):467-473.,
[5] Zhao Y, Huang Q, Zhang L, et al. Squeeze Film Air Damping in Tapping Mode Atomic Force Microscopy[J]. Micromachines, 2017, 8(7):226.
[6] Hoffmann P M, Jeffery S, Pethica J B, et al. Energy dissipation in atomic force microscopy and atomic loss processes[J]. Physical Review Letters, 2001, 87(26):265502.
[7] Bao M, Yang H. Squeeze film air damping in MEMS[J]. Sensors & Actuators A Physical, 2007, 136(1):3-27.
[8] 冯闯,赵亚溥,刘冬青. 微梁谐振器中的空气阻尼[J]. 工程科学学报, 2007, 29(8):841-844.
[9] Kavitha C, Madhan M G. Study of squeeze film damping characteristics under different gas mediums in a capacitive MEMS accelerometer[J]. Journal of the Brazilian Society of Mechanical Sciences & Engineering, 2016, 38(1):241-252.
[10] Starr J B . Squeeze-film damping in solid-state accelerometers[C]// IEEE Solid-state Sensor & Actuator Workshop. IEEE, 1990.
[11] 张文明, 孟光, 周健斌. 微机电系统压膜阻尼特性分析[J]. 振动与冲击, 2006, 25(4):41-45.
[12] Young Y J. Squeeze-film damping for MEMS structures[J]. 1998.
[13] Hosaka H, Itao K, Kuroda S. Damping characteristics of beam-shaped micro-oscillators[J]. Sensors & Actuators A Physical, 1995, 49(1):87-95.
[14] 李普,胡如夫,尹垚. 弹性悬臂微梁谐振系统挤压膜阻尼新解析模型[J]. 振动与冲击, 2008, 27(3):1-4.
[15] Naik T, Longmire E K, Mantell S C. Dynamic response of a cantilever in liquid near a solid wall[J]. Sensors & Actuators A Physical, 2003, 102(3):240-254.
[16] Xu Y, Smith S T. Determination of squeeze film damping in capacitance-based cantilever force probes[J]. Precision Engineering, 1995, 17(2):94-100.
[17] Rankl C, Pastushenko V, Kienberger F, et al. Hydrodynamic damping of a magnetically oscillated cantilever close to a surface[J]. Ultramicroscopy, 2004, 100(3):301-308.
[18] Korayem M H, Ebrahimi N. Nonlinear dynamics of tapping-mode atomic force microscopy in liquid[J]. Journal of Applied Physics, 2011, 109(8):064107.
[19] Korayem M H, Ebrahimi N, Korayem A H. Modeling and Simulation of Tapping-mode Atomic Force Microscopy in Liquid[J]. Nanoscience & Nanotechnology, 2011, 1(1):14-21.
[20] 刘鸿文. 高等材料力学[M]. 高等教育出版社, 1985.
[21] Lévêque G, Girard P, Belaidi S, et al. Effects of air damping in noncontact resonant force microscopy[J]. Review of Scientific Instruments, 1997, 68(11):4137-4144.
[22] 鲍敏杭,孙远程,杨恒,等. 微电子机械器件压膜空气阻尼的一般化雷诺方程[J]. 半导体学报, 2002, 23(12):1245-1248.
[23] Chan D Y C, Horn R G. The drainage of thin liquid films between solid surfaces[J]. Journal of Chemical Physics, 1985, 83(10):5311-5324.
[24] Elsharkawy A A, Al-Fadhalah K J. Separation of a sphere from a flat in the presence of couple stress fluids[J]. Lubrication Science, 2010, 20(1):61-74.
[25] 丁文璇,刘运鸿,魏征. 原子力显微镜探针振动的简化模型分析[J]. 力学与实践, 2017, 39(5):449-454.
[26] Garcỉa A R , Rubén Pérez. Dynamic atomic force microscopy methods[J]. Surface Science Reports, 2002, 47(6-8):197-301.
[27] 魏征,赵爽,陈少勇,等. 原子力显微镜中液桥生成机理探讨[J]. 应用数学和力学, 2015, 36(1):87-98.