中国计量科学研究院研制出频率范围0.0005Hz~1200Hz高精度角振动基准装置,低频角振动装置最大角位移300°,中频角振动装置最大角位移60°,角加速度范围从0.04 rad/s2到2000 rad/s2。装置实现了衍射式外差激光干涉仪测量方法,以及圆光栅和双光束外差激光干涉仪差动测量。角加速度复灵敏度测量不确定度(k=2):参考点优于0.5%,0.5°,通频带优于1.0%,1°。
Abstract
Primary standard of angular vibration devices have been developed in the National institute of Metrology. The maximum angular displacement of low angular vibration device is 300 ° and medium angular vibration device is 60 °. The dynamic range of angular accelerate is 0.04 rad/s2 to 2 000 rad/s2. Diffraction grating and heterodyne laser interferometer acted as the main measurement method. At the same time, circular grating and twobeam heterodyne laser interferometer differential measurement have been realized. Uncertainty of measurement of angular accelerate sensitivity (k=2) is: 0.5%, 0.5° (reference), and others are better than 1.0%, 1.0° in whole frequency band.
{{custom_sec.title}}
{{custom_sec.title}}
{{custom_sec.content}}
参考文献
[1] ISO 16063-15:2006, Methods for the calibration of vibration and shock transducers - Part 15: Primary angular vibration calibration by laser interferometry [S].
[2] Angelika Täubner, Hans-Jürgen von Martens. Measurement of angular accelerations, angular velocities and rotation angles by grating interferometry [J]. Measurement 24(1998) 21-32;
[3] Hans-Jürgen von Martens, Angelika Täubner, Wolfgang Wabinski, Alfred Link, Hans-Joachim Schlaak. Traceability of vibration and shock measurements by laser interferometry[J]. Measurement 24(1998) 21-32;
[4] Angelika Täubner, Hans-Jürgen von Martens. Development and investigation of diffraction grating interferometers, Proceedings First International Conference on Vibration Measurement by Laser Technique, Ancona, Italy, 1994, p.145;
[5] Wan-Sup Cheung and Cheol-Ung cheong, Angle Prism-Based laser interferometer for high prision measurement of angular vibration, IMEKO XVIII World Congress, Rio de Jan eiro,Brasil,2006;
[6] Wan-Sup Cheung and sang-Myong Park, Progress in development of primary angular vibration calibration systems, TC3, TC16 & TC22 IMEKO international conference, in Merida, Mexico, 2007;
[7] 彭军,何群,薛景峰,孙浩. 低频标准角振动台[J]. 测量与校准,2005,25(6),46-48;
PENG Jun, HE Qun, Xue Jing-feng, SUN Hao. Low frequency standard angular exciter, Metrology & Measurement technology[J], 2005,25(6),46-48;
[8] 薛景峰,彭军,李新良,赵维谦. 角振动校准装置研究[J]. 航空学报,2015,36(3);
XUE Jing-feng, PENG Jun, LI Xinliang, ZHAO Weiqian. Research on angular vibration calibration device[J]. Acta Aeronautica et Astronautica Sinica. 2015,36(3);
[9] 于梅,刘爱东等.衍射光栅外差激光干涉法角振动校准技术研究[J]. 计量学报,2015,36(6),561-564.
YU Mei, LIU Ai-dong, HE Wen, Yang Li-feng, ZUO Ai-bin, Hu Hong-bo. Angle vibration calibration technology by diffraction grating heterodyne laser interferometry[J]. Acta Metrologica Sinica, 2015,36(6),561-564.
[10] 刘爱东,于梅. 漫反射式零差正交激光测振仪[J].振动与冲击,2015,34(23),212-215。
Liu Ai-dong, YU Mei. Diffuse reflection type homodyne quadrature laser vibrometers[J]. Journal of Vibration and Shock, 2015,34(23),212-215.