Abstract

Primary standard of angular vibration devices have been developed in the National institute of Metrology. The maximum angular displacement of low angular vibration device is 300 ° and medium angular vibration device is 60 °. The dynamic range of angular accelerate is 0.04 rad/s2 to 2 000 rad/s2. Diffraction grating and heterodyne laser interferometer acted as the main measurement method. At the same time, circular grating and twobeam heterodyne laser interferometer differential measurement have been realized. Uncertainty of measurement of angular accelerate sensitivity (k=2) is: 0.5%,  0.5° (reference), and others are better than 1.0%, 1.0° in whole frequency band.

Cite this article

Download Citations
LIU Aidong1, YU Mei1, HE Wen2. A study of an angle vibration measuring method[J]. Journal of Vibration and Shock, 2018, 37(12): 216-219

References

[1] ISO 16063-15:2006, Methods for the calibration of vibration and shock transducers - Part 15: Primary angular vibration calibration by laser interferometry [S].
[2] Angelika Täubner, Hans-Jürgen von Martens. Measurement of angular accelerations, angular velocities and rotation angles by grating interferometry [J]. Measurement 24(1998) 21-32;
[3] Hans-Jürgen von Martens, Angelika Täubner, Wolfgang Wabinski, Alfred Link, Hans-Joachim Schlaak. Traceability of vibration and shock measurements by laser interferometry[J]. Measurement 24(1998) 21-32;
[4] Angelika Täubner, Hans-Jürgen von Martens. Development and investigation of diffraction grating interferometers, Proceedings First International Conference on Vibration Measurement by Laser Technique, Ancona, Italy, 1994, p.145;
[5] Wan-Sup Cheung and Cheol-Ung cheong, Angle Prism-Based laser interferometer for high prision measurement of angular vibration, IMEKO XVIII World Congress, Rio de Jan eiro,Brasil,2006;
[6]   Wan-Sup Cheung and sang-Myong Park, Progress in development of primary angular vibration calibration systems, TC3, TC16 & TC22 IMEKO international conference, in Merida, Mexico, 2007;
[7] 彭军,何群,薛景峰,孙浩. 低频标准角振动台[J]. 测量与校准,2005,25(6),46-48;
   PENG Jun, HE Qun, Xue Jing-feng, SUN Hao. Low frequency standard angular exciter, Metrology & Measurement technology[J], 2005,25(6),46-48;
[8] 薛景峰,彭军,李新良,赵维谦. 角振动校准装置研究[J]. 航空学报,2015,36(3);
   XUE Jing-feng, PENG Jun, LI Xinliang, ZHAO Weiqian. Research on angular vibration calibration device[J]. Acta Aeronautica et Astronautica Sinica. 2015,36(3);
[9] 于梅,刘爱东等.衍射光栅外差激光干涉法角振动校准技术研究[J]. 计量学报,2015,36(6),561-564.
   YU Mei, LIU Ai-dong, HE Wen, Yang Li-feng, ZUO Ai-bin, Hu Hong-bo. Angle vibration calibration technology by diffraction grating heterodyne laser interferometry[J]. Acta Metrologica Sinica, 2015,36(6),561-564.
[10] 刘爱东,于梅. 漫反射式零差正交激光测振仪[J].振动与冲击,2015,34(23),212-215。
    Liu Ai-dong, YU Mei. Diffuse reflection type homodyne quadrature laser vibrometers[J]. Journal of Vibration and Shock, 2015,34(23),212-215.